Frigga M2D

Frigga M2D Reusable USB Temperature Data Logger

Model: M2D

Zavedení

The Frigga M2D is a reusable USB temperature data logger designed for accurate and reliable temperature monitoring. It features an advanced internal sensor with a measuring range of -30°C to +70°C (-22°F to +158°F) and a memory capacity of up to 14,400 readings. This device is suitable for various applications requiring precise temperature recording, such as cold chain logistics, food storage, pharmaceutical monitoring, and laboratory use.

Nastavení

Before initial use, the Frigga M2D data logger requires configuration using the complimentary Frigga Data Software. This software is compatible with Windows and Mac OS and allows you to set logging conditions such as delayed start, logging intervals, temperature units (Fahrenheit/Celsius), time zones, and temperature alarm ranges. The software can be downloaded from the official Frigga webmísto.

Important Configuration Notes:

  • Ensure the 'Stop' button function is set to 'Allow' in the software settings.
  • Ensure the 'Restart' option is set to 'Allow' for subsequent uses.
  • The device is factory-set with a 30minutové zpoždění startu. During this delay, the logger will not immediately begin recording after pressing the start button. An hourglass icon will appear on the display, and the blue light will flash every 10 seconds. This delay allows the logger to acclimate to its environment before recording begins, preventing initial out-of-range readings.
Frigga M2D data logger connected to a computer, showing software interface for parameter settings.

Image: The Frigga M2D data logger connected to a computer, illustrating the software interface used for setting various parameters like temperature units, alarm ranges, and logging intervals.

Návod k obsluze

Spuštění a zastavení nahrávání

  • Na start recording, press the 'START/STATUS' button. The device will begin logging data after any configured start delay.
  • Na zastávka recording, press the 'STOP' button.

ViewData v reálném čase

The integrated LCD screen displays real-time temperature readings, alarm status, and statistical information. You can switch between Fahrenheit and Celsius units directly on the device.

LED indikátor alarmu

The device features an LED alarm function that activates if the recorded temperature exceeds or falls below the pre-set thresholds configured in the software.

Načítání dat

After recording is complete, plug the USB connector of the data logger into any computer. The device will automatically generate PDF and CSV reports, which can be accessed directly without requiring additional software or driver installation. The PDF report provides a summary and graph, while the CSV file contains raw data points.

Frigga M2D data logger generating PDF and CSV reports when plugged into a computer.

Image: The Frigga M2D data logger connected via USB to a computer, demonstrating its ability to auto-generate PDF and CSV reports for data analysis.

Údržba

Výměna baterie

The Frigga M2D uses a CR2450 battery. When the battery indicator shows low power, replace the battery to ensure continuous operation and data integrity. Follow these steps:

  1. Najděte přihrádku na baterie na zadní straně zařízení.
  2. Use a coin or a flat tool to twist the battery cover counter-clockwise to open.
  3. Vyjměte starou baterii CR2450.
  4. Vložte novou baterii CR2450 kladnou (+) stranou nahoru.
  5. Replace the battery cover and twist clockwise to secure.
Four-step visual guide on how to replace the CR2450 battery in the Frigga M2D data logger.

Image: A step-by-step visual guide demonstrating the process of replacing the CR2450 battery in the Frigga M2D data logger.

Čištění

Otřete zařízení dočista měkkým hadříkemamp cloth. Do not use abrasive cleaners or immerse the device in water, despite its IP65 rating, to prevent damage to the USB port or internal components.

Odstraňování problémů

  • Device does not start recording immediately: The device has a factory-set 30-minute start delay. Wait for this period to elapse. An hourglass icon and a flashing blue light indicate the delay is active. Ensure the 'Start Delay' setting in the Frigga Data Software is configured as desired.
  • Software issues (e.g., blank window, crashes): If the Frigga Data Software does not open correctly (e.g., a blank window appears), try closing the process via Task Manager and restarting the software. Ensure your operating system (Windows/Mac OS) is updated and meets the software's requirements.
  • Inaccurate or unexpected readings: Ensure the device has acclimated to the environment before starting a recording session. Verify that the temperature alarm ranges and logging intervals are correctly set in the Frigga Data Software.
  • Zařízení nebylo počítačem rozpoznáno: Ensure the USB connection is secure. Try plugging the device into a different USB port or another computer.
  • Krátká výdrž baterie: The initial battery might have a reduced charge. Replace the CR2450 battery with a new one if you experience unusually short operational periods.

If you encounter persistent issues, please refer to the support section for contact information.

Specifikace

ModelM2D
Rozsah měření-30 °C až +70 °C (-22 °F až +158 °F)
Přesnost±0.5°C (-20°C to +40°C); ±1°C for other ranges
Kapacita paměti14,400 přečtení
BaterieCR2450 (vyměnitelný)
KonektivitaUSB
Typ zobrazeníDigitální LCD
Vodotěsná třídaIP65
Rozměry (přibližně)11.2 cm (4.41 in) Height x 4.7 cm (1.85 in) Width
Speciální funkceAlarm, Fast Reading System, High Accuracy, Low Battery Indicator
Diagram showing the Frigga M2D data logger with its key product parameters including dimensions, measurement range, accuracy, and memory capacity.

Image: A visual representation of the Frigga M2D data logger, highlighting its physical dimensions and core technical specifications.

Záruka a podpora

The Frigga M2D data logger comes with a warranty. Details regarding the warranty period and terms can be found on the warranty card included with your product. You can also scan the QR code on the warranty card for access to the user manual and additional resources.

For technical assistance or any inquiries, 24/7 technician support is available via email. Please refer to the contact information provided on the warranty card or the official Frigga webmísto.

A calibration certificate can be exported by connecting the device to your computer.

Související dokumenty - M2D

Předview Frigga T7 Series User Manual - Real-Time Cold Chain Temperature Monitoring
Comprehensive user manual for the Frigga T7 Series temperature data logger. Covers device appearance, display functions, setup, monitoring, stopping, reporting, and important cautions for cold chain logistics.
Předview Uživatelská příručka k záznamníku teploty a vlhkosti FRIGGA V5
Uživatelská příručka k záznamníku teploty a vlhkosti FRIGGA V5 s podrobným popisem jeho funkcí, obsluhy, nastavení a reportovacích funkcí pro logistiku chladírenského řetězce.
Předview Uživatelská příručka k záznamníku teploty a vlhkosti FRIGGA řady V5 Plus
Komplexní uživatelská příručka k záznamníku teploty a vlhkosti FRIGGA řady V5 Plus. Naučte se, jak nastavit, obsluhovat a načítat data ze záznamníku pro monitorování chladicího řetězce.
Předview Uživatelská příručka k multisenzorovému senzoru Frigga 4G a Wi-Fi 6 v reálném čase
Uživatelská příručka pro multisenzor Frigga 4G & Wi-Fi 6 v reálném čase (model S0) s podrobným popisem jeho funkcí, obsahu balení a informací o produktu.view, pokyny k nastavení, průvodce LED indikátory, technické specifikace, umístění zařízení, informace o baterii, kroky pro řešení problémů a shoda s FCC.
Předview Uživatelská příručka k datovému centru Frigga V 1.0.0
Komplexní uživatelská příručka k softwaru Frigga Data Center, která uživatele provede instalací, správou dat, konfigurací zařízení, cloudovými službami a funkcemi pro monitorování chladicího řetězce.
Předview Frigga SV2 Series Smart Active Data Logger User Manual
User manual for the Frigga SV2 Series Smart Active Data Logger, detailing appearance, display functions, operation steps, usage scenarios, and FCC compliance.