Amprobe SM-CAL1

AmpUživatelská příručka ke kalibrátoru hlukoměru robe SM-CAL1

Model: SM-CAL1

1. Úvod

The Amprobe SM-CAL1 is a precision sound meter calibrator designed to ensure the accuracy and reliability of your sound level measurements. This device generates two fixed sound pressure levels, 94 decibels (dB) and 114 dB, at a frequency of 1,000 hertz (Hz), allowing for precise calibration of sound meters with microphones up to 1/2 inch in diameter. Regular calibration is crucial for maintaining measurement integrity and compliance with relevant standards.

This manual provides detailed instructions for the proper use, maintenance, and troubleshooting of your SM-CAL1 Sound Meter Calibrator.

2. Bezpečnostní informace

Please read and understand all safety instructions before operating the SM-CAL1. Failure to follow these guidelines may result in injury or damage to the device.

  • Do not expose the calibrator to extreme temperatures, humidity, or direct sunlight.
  • Zabraňte pádu zařízení nebo jeho vystavení silným nárazům.
  • Do not attempt to open or modify the calibrator. Refer all servicing to qualified personnel.
  • Během provozu se ujistěte, že je přihrádka na baterie bezpečně uzavřena.
  • Zlikvidujte baterie v souladu s místními předpisy.

3. Obsah balení

Ověřte, zda jsou v balení přítomny všechny položky:

  • Amprobe SM-CAL1 Sound Meter Calibrator
  • One (1) 9V battery (installed)
  • Pokyny pro uživatele (tato příručka)

4. Nastavení

4.1 Instalace baterie

The SM-CAL1 comes with a 9V battery pre-installed. If replacement is needed:

  1. Vyhledejte kryt přihrádky na baterie na spodní straně jednotky.
  2. Odstraňte kryt.
  3. Connect a new 9V battery, observing correct polarity.
  4. Bezpečně nasaďte kryt prostoru pro baterie.

4.2 Preparing for Calibration

Before calibrating your sound meter, ensure its microphone is clean and free of debris. The SM-CAL1 is designed to fit microphones with a diameter of up to 1/2 inch.

Amprobe SM-CAL1 Sound Meter Calibrator with power switch and display showing 94 114 OFF db at 1kHz, IEC 942 1988 CLASS 2, and Sound Calibrator.

Obrázek 1: Amprobe SM-CAL1 Sound Meter Calibrator. The image shows the top face of the cylindrical calibrator, featuring the 'AMPROBE' logo, 'SM-CAL1' model number, a three-position slide switch labeled '94', 'OFF', and '114' (referring to dB levels), and text indicating 'db at 1kHz', 'IEC 942 1988 CLASS 2', and 'Sound Calibrator'. A red power indicator light is also visible.

5. Provoz

5.1 Calibrating a Sound Meter

  1. Gently insert the microphone of your sound level meter into the opening on the top of the SM-CAL1 calibrator. Ensure a snug fit to minimize sound leakage.
  2. Select the desired calibration level using the slide switch on the calibrator:
    • Slide to '94' for a 94 dB signal.
    • Slide to '114' for a 114 dB signal.
  3. Observe the reading on your sound level meter. Adjust your sound meter's calibration settings until its display matches the selected output level of the SM-CAL1 (either 94 dB or 114 dB). Refer to your sound meter's user manual for specific calibration adjustment procedures.
  4. Once calibration is complete, slide the switch to 'OFF' to conserve battery life.
  5. Carefully remove the sound meter's microphone from the calibrator.

5.2 Indikace slabé baterie

The SM-CAL1 is equipped with a low-battery indicator. When the battery voltage drops below a certain level, the indicator light will dim or not illuminate, signaling that the 9V battery needs to be replaced.

6. Údržba

6.1 Čištění

To clean the exterior of the calibrator, use a soft, dry cloth. Do not use abrasive cleaners or solvents, as these may damage the casing nebo vnitřních součástí.

6.2 Skladování

When not in use, store the SM-CAL1 in a cool, dry place, away from direct sunlight and extreme temperatures. If storing for extended periods, it is recommended to remove the 9V battery to prevent leakage.

6.3 Recalibration of the SM-CAL1

While the SM-CAL1 is designed for stable output, periodic verification of its accuracy by a certified calibration laboratory is recommended to ensure continued compliance with standards and optimal performance. The frequency of such recalibration depends on usage and regulatory requirements.

7. Řešení problémů

ProblémMožná příčinaŘešení
No sound output / Power indicator offBaterie je vybitá nebo nesprávně nainstalovaná.Vyměňte 9V baterii a dbejte na správnou polaritu.
Inaccurate calibration readings on sound meterMicrophone not properly seated; Calibrator needs recalibration; Sound meter issue.Ensure microphone is snugly fitted. Consider sending SM-CAL1 for professional recalibration. Check your sound meter's functionality.
Sound meter reading fluctuatesAmbient noise interference; Loose microphone connection.Perform calibration in a quiet environment. Re-seat the microphone firmly.

8. Specifikace

  • Model: SM-CAL1
  • Output Sound Pressure Levels: 94 dB a 114 dB
  • Výstupní frekvence: 1,000 Hz (1 kHz)
  • Kompatibilita mikrofonu: Fits microphones up to 1/2 inch diameter
  • Přesnost: Meets IEC 942 1988 Class 2 standards
  • Zdroj napájení: One (1) 9V battery
  • Indikátor slabé baterie: Ano
  • Provozní teplota: 0 °C až 40 °C (32 °F až 104 °F)
  • Skladovací teplota: -10 °C až 50 °C (14 °F až 122 °F)
  • Dimensions (H x D): Přibližně 125 x 48 mm (4.9 x 1.9 palce)
  • Hmotnost: Přibližně 0.6 lb (285 g)
  • certifikace: CE compliant, ANSI, IEC standards

9. Záruka a podpora

9.1 Informace o záruce

Amprobe products are warranted against defects in material and workmanship for a specified period from the date of purchase. For detailed warranty terms and conditions, please refer to the warranty card included with your product or visit the official Ampžupan website. This warranty does not cover damage resulting from misuse, accident, neglect, alteration, contamination, or abnormal conditions of operation or handling.

9.2 Zákaznická podpora

Pokud narazíte na jakékoli problémy nebo máte otázky týkající se vašeho Amprobe SM-CAL1 Sound Meter Calibrator, please contact Amprobe customer support. Before contacting support, ensure you have reviewed the troubleshooting section of this manual. You can find contact information on the official Ampžupan webna webu nebo prostřednictvím místního distributora.

Související dokumenty - SM-CAL1

Předview AmpUživatelská příručka digitálního multimetru robe AM-420
Uživatelská příručka pro AmpDigitální multimetr robe AM-420 s podrobným popisem jeho funkcí, měřicích možností, bezpečnostních opatření a specifikací pro elektrické testování. Vhodný pro domácí i profesionální použití.
Předview AmpUživatelská příručka k měřiči kapacity a odporu robe CR50A
Tato uživatelská příručka poskytuje komplexní pokyny k obsluze AmpMěřič kapacity a odporu robe CR50A, který zahrnuje jeho funkce, postupy měření, bezpečnostní pokyny a technické specifikace.
Předview AmpUživatelská příručka k systému pro vyhledávání podzemních kabelů/potrubí robe AT-3500
Komplexní uživatelská příručka pro AmpSystém pro vyhledávání podzemních kabelů/potrubí robe AT-3500, podrobný popis provozu, funkcí, technických specifikací a údržby pro přesnou detekci podzemních inženýrských sítí.
Předview Amprobe TT-200 Butt Set Instruction Manual - Line Testing and Communications
Komplexní návod k obsluze pro Amprobe TT-200 Butt Set, detailing its features for line testing, insulation resistance measurement, and temporary communications for technicians.
Předview AmpUživatelská příručka k infračervenému teploměru robe IR-750 / IR-750-EUR 50:1
Uživatelská příručka pro AmpInfračervené teploměry robe IR-750 a IR-750-EUR 50:1. Mezi jejich vlastnosti patří široký teplotní rozsah, vysoká přesnost, laserové ukazovátko, záznam dat a USB připojení pro průmyslové a profesionální aplikace.
Předview AmpVyhledávač podzemních inženýrských sítí robe UAT-600 – stručný návod k obsluze
Stručný úvodní průvodce pro AmpLokátor podzemních inženýrských sítí robe UAT-600, který poskytuje pokyny k aktivní a pasivní lokalizaci inženýrských sítí s využitím různých režimů, jako je napájení, rádio, indukce, přímé testovací kabely a signální cívka.amp.