1. Úvod
This manual provides detailed instructions for the safe and effective operation of the WOOXGEHM T6125 Digital Insulation Resistance Tester. The T6125 is a robust, shockproof, and dustproof instrument designed for measuring insulation resistance, absorption ratio (DAR), polarization index (PI), AC voltage, and low resistance. It features a wide measurement range, multiple output voltage levels, a large LCD with backlight, and strong anti-interference capabilities, making it suitable for various electrical equipment testing applications.

Figure 1: WOOXGEHM T6125 Digital Insulation Resistance Tester.
2. Bezpečnostní informace
To ensure safe operation and prevent damage to the instrument or injury to personnel, please read and understand all safety instructions before use.
- Vždy dodržujte místní a národní bezpečnostní předpisy.
- Do not operate the tester if it appears damaged or is not functioning properly.
- Ensure the circuit under test is de-energized and discharged before connecting test leads for resistance measurements.
- Používejte pouze dodané měřicí kabely a příslušenství.
- Během měření se nedotýkejte měřicích vodičů ani měřeného obvodu, zejména při vysokém napětí.tages jsou přítomny.
- Do not use the instrument in wet environments or explosive atmospheres.
- Jakmile se zobrazí indikátor slabých baterií, ihned vyměňte baterie.
3. Obsah balení
Ověřte, zda jsou v balení přítomny všechny níže uvedené položky:
- 1 x WOOXGEHM T6125 Digital Insulation Resistance Tester
- 1 x pár testovacích vodičů
- 1 x External Power Plug (5.5mm)
- 2 x krokosvorky
- 1 x látkový vak
- 1 x Uživatelská příručka (anglicky)
4. Konec produktuview
The T6125 insulation tester is designed for comprehensive electrical insulation testing. Key features include:
- Měření izolačního odporu: Measures insulation opposition, absorption ratio, and polarization index.
- AC svtage Měření: Schopný měřit střídavý proud (AC Volt)tage.
- Měření nízkého odporu: For continuity checks.
- Multiple Test Voltages: Offers 100V, 250V, 500V, 1000V, and 2500V test voltage možnosti.
- Funkce automatického vybíjení: Zajišťuje bezpečnost po vysokém objemutage testy.
- Pozastavení dat: Zmrazí zobrazené měření pro snadné zaznamenání.
- Over-range Alarm: Alerts the user when measurements exceed the selected range.
- Automatické vypnutí: Vypne se po 30 minutách nečinnosti, aby se šetřila baterie.
- Velký podsvícený LCD displej: Poskytuje jasné odečty za různých světelných podmínek.
- Odolný design: Shockproof, dustproof, and dampness-proof construction.

Figure 2: Front panel layout and key components of the T6125.
- Vysoký objemtage Output Black Negative Jack: Connects the negative test lead for high voltage testy.
- Vysoký objemtage Output Red Positive Jack: Connects the positive test lead for high voltage testy.
- Resistor Red Positive Jack: Connects the positive test lead for resistance measurements.
- Resistive Black Negative Jack: Connects the negative test lead for resistance measurements.
- 12V Power Input Jack: Pro externí napájení.
- LCD displej: Zobrazuje naměřené hodnoty a stav.
- Testovací tlačítko: Zahájí měření.
- Tlačítko pro pozastavení dat: Zmrazí aktuální zobrazení.
- Tlačítko podsvícení: Activates/deactivates the LCD backlight.
- DAR (Absorption Ratio) / PI (Polarization Index) Button: Selects DAR or PI measurement mode.
- Gear Rotary Switch: Selects measurement function and test voltage.

Figure 3: The T6125 features an automatic shutdown function after 30 minutes of inactivity to conserve battery life.
5. Nastavení
5.1 Instalace baterie
The T6125 requires 8 x 1.5V AA batteries (not included).
- Najděte přihrádku na baterie na zadní straně jednotky.
- Otevřete kryt bateriového prostoru.
- Vložte 8 baterie AA, dbejte na správnou polaritu (+/-).
- Bezpečně zavřete kryt bateriového prostoru.
5.2 Připojení měřicích kabelů
Connect the test leads to the appropriate jacks on the front panel based on the measurement type. For insulation resistance tests, use the High Voltage Output jacks. For low resistance, use the Resistor jacks.
6. Návod k obsluze
6.1 Zapnutí/vypnutí
Rotate the Gear Rotary Switch from the "OFF" position to any measurement function to power on the device. Rotate back to "OFF" to power off. The device will automatically shut down after 30 minutes of inactivity.
6.2 Měření izolačního odporu
This function measures the insulation resistance of electrical equipment.
- Ujistěte se, že testovaný obvod je bez napětí a bezpečně vybitý.
- Connect the test leads to the High Voltage Output jacks (red to positive, black to negative).
- Connect the other ends of the test leads to the insulation points of the equipment to be tested.
- Rotate the Gear Rotary Switch to the desired insulation test voltage (např. 100 V, 250 V, 500 V, 1000 V, 2500 V).
- Stiskněte a podržte TEST button. The measurement will begin, and the resistance value will be displayed on the LCD.
- Uvolněte TEST button to stop the measurement. The instrument will automatically discharge the circuit.
- To hold the reading, press the DRŽET button during or after a measurement. Press again to release.

Figure 4: The T6125 offers a wide insulation test measurement range up to 2500V and 200.0GΩ.
6.3 Měření absorpčního poměru (DAR) a polarizačního indexu (PI)
DAR and PI are used to assess the quality of insulation over time.
- Follow steps 1-3 for Insulation Resistance Measurement.
- Rotate the Gear Rotary Switch to the desired insulation test voltage.
- Stiskněte tlačítko DAR/PI button to select the desired mode. The display will indicate "DAR" or "PI".
- Stiskněte a podržte TEST button. The instrument will automatically perform the timed measurements and display the DAR or PI value.
- Uvolněte TEST tlačítko.
6.4 AC svtage Měření
Pro měření AC objtage:
- Rotate the Gear Rotary Switch to the "ACV" position.
- Připojte měřicí vodiče k obvodu, kde je střídavý proud.tagJe třeba měřit e.
- AC svtagNa LCD se zobrazí e.
6.5 Měření nízkého odporu
For continuity or low resistance checks:
- Rotate the Gear Rotary Switch to the "Ω" (low resistance) position.
- Connect the test leads to the Resistor jacks.
- Connect the other ends of the test leads to the component or circuit to be measured.
- The resistance value will be displayed on the LCD.
7. Údržba
7.1 Čištění
Otřete nástroj reklamouamp cloth and mild detergent. Do not use abrasive cleaners or solvents. Ensure the instrument is completely dry before storage or next use.
7.2 Skladování
When not in use for extended periods, remove the batteries to prevent leakage. Store the instrument in the provided cloth bag in a cool, dry place, away from direct sunlight and extreme temperatures.
- Skladovací teplota: -20 ℃ až 60 ℃
- Working Conditions: 0℃ to 40℃
- Relativní vlhkost: 40% až 75%
8. Řešení problémů
If you encounter issues with your WOOXGEHM T6125, refer to the following common problems and solutions:
- Bez napájení:
- Zkontrolujte, zda jsou baterie správně vloženy a zda jsou dostatečně nabité. V případě potřeby je vyměňte.
- Ensure the Gear Rotary Switch is not in the "OFF" position.
- Nepřesná čtení:
- Verify test leads are securely connected and not damaged.
- Ensure the correct measurement function and test voltage are selected.
- Clean the test points on the equipment being measured.
- Zkontrolujte vnější zdroje rušení.
- "OL" or Over-range Display:
- The measured value exceeds the selected range. Try a higher test voltage or a different range if applicable.
- Zajistěte správné připojení k obvodu.
Pokud problém přetrvává, kontaktujte zákaznickou podporu.
9. Specifikace
| Model | T6125 |
| Materiál | ABS |
| Napájení | 8 x 1.5V AA baterie (nejsou součástí balení) |
| Velikost položky | 179 x 133 x 70 mm (7.04 x 5.23 x 2.75 palce) |
| Hmotnost položky | 570 g (1.25 lb) |
| Skladovací teplota | -20 ℃ až 60 ℃ |
| Pracovní podmínky | 0℃ až 40℃ |
| Nadmořská výška | 2000 m |
| Relativní vlhkost | 40 % až 75 % |
Insulation Test Accuracy (T-6125)
| Výstupní objemtage | Rozsah | Rezoluce | Maximální zkušební proud | Přesnost |
|---|---|---|---|---|
| 250V (± 10%) | 0~20MΩ | 0.01 MΩ | 2 mA | ± (3 % + 5) |
| 20~200MΩ | 0.1 MΩ | ± (3 % + 5) | ||
| 0.2G~2GΩ | 0.001 GΩ | ± (10 % + 5) | ||
| 500V (± 10%) | 0~20MΩ | 0.01 MΩ | 2 mA | ± (3 % + 5) |
| 20~200MΩ | 0.1 MΩ | ± (3 % + 5) | ||
| 0.2G~2GΩ | 0.001 GΩ | ± (10 % + 5) | ||
| 1000V (± 10%) | 0~20MΩ | 0.01 MΩ | 2 mA | ± (3 % + 5) |
| 20~200MΩ | 0.1 MΩ | ± (3 % + 5) | ||
| 0.2G~2GΩ | 0.001 GΩ | ± (10 % + 5) | ||
| 2500V (± 10%) | 0~20MΩ | 0.01 MΩ | 2 mA | ± (3 % + 5) |
| 20~200MΩ | 0.1 MΩ | ± (5 % + 5) | ||
| 0.2G~2GΩ | 0.001 GΩ | ± (10 % + 5) | ||
| 2G~20GΩ | 0.01 GΩ | ± (10 % + 5) | ||
| AC svtage Přesnost | Ranging: 600V | Rozlišení: 1V | N/A | ± (1.5 % + 5) |
| Přesnost odporu | Ranging: 60Ω | Rozlišení: 0.01Ω | N/A | ± (1.2 % + 5) |

Figure 5: Detailed insulation test accuracy specifications for the T6125 model.

Figure 6: Physical dimensions of the WOOXGEHM T6125.
10. Záruka a podpora
For warranty information or technical support, please refer to the contact details provided with your purchase or visit the official WOOXGEHM webwebu. Uschovejte si účtenku jako doklad o koupi.