U.S. Solid USS-DBS64-500

U.S. Solid 500g x 0.001g Lab Analytical Balance Instruction Manual

Model: USS-DBS64-500

1. Úvod

This manual provides essential instructions for the proper setup, operation, maintenance, and troubleshooting of your U.S. Solid 500g x 0.001g Lab Analytical Balance. Designed for high-precision weighing, this balance offers reliable results with 0.001g (1mg) accuracy and a maximum capacity of 500g. It features multiple unit conversions, parts counting, percentage weighing, and manual calibration capabilities.

Please read this manual thoroughly before using the balance to ensure accurate measurements and safe operation.

2. Bezpečnostní opatření

  • Always operate the balance on a stable, level, and vibration-free surface.
  • Avoid placing the balance near strong electromagnetic fields, direct sunlight, or rapid temperature changes.
  • Nepřetěžujte váhu nad její maximální nosnost 500 g.
  • Keep the balance and its components clean and free from dust and corrosive substances.
  • Use only the provided power adapter or a compatible 9V battery.
  • Do not attempt to disassemble or repair the balance yourself. Contact qualified service personnel if needed.

3. Co je v krabici

Carefully unpack all components and check against the following list:

  • Scale Body x1
  • Scale Pan x1
  • Napájecí kabel 1x
  • Calibration Weight x1
  • Návod k použití x1

4. Konec produktuview

Familiarize yourself with the main components of your analytical balance.

Přední view of the U.S. Solid Lab Analytical Balance with windshield and display showing 0.000g.
Obrázek 4.1: Přední view of the balance. Shows the main display, control buttons, weighing pan, and windshield.
Strana view of the U.S. Solid Lab Analytical Balance, highlighting the compact design.
Obrázek 4.2: Strana view of the balance, illustrating its compact form factor.
Zadní view of the U.S. Solid Lab Analytical Balance, showing the power input.
Obrázek 4.3: Zadní view of the balance, including the power input port.
Close-up of the bubble level on the U.S. Solid Lab Analytical Balance for horizontal adjustment.
Obrázek 4.4: Detail of the integrated bubble level for precise horizontal adjustment.
Explodoval view of the U.S. Solid Lab Analytical Balance components with dimensions.
Obrázek 4.5: Explodoval view showing all components and their dimensions.

5. Nastavení

5.1 Vybalení a počáteční kontrola

Carefully remove the balance from its packaging. Inspect all components for any signs of damage. If any damage is found, do not use the balance and contact customer support immediately.

5.2 Umístění

Place the balance on a firm, level, and stable surface. Ensure the location is free from vibrations, drafts, direct sunlight, and rapid temperature fluctuations, which can affect measurement accuracy.

5.3 Vyrovnání váhy

The balance must be perfectly level for accurate readings. Use the adjustable feet at the bottom of the balance to center the bubble in the integrated level indicator (refer to Figure 4.4).

Image showing the horizontal adjustment process with the bubble level.
Obrázek 5.1: Horizontal Adjustment. Adjust leveling feet until the bubble is centered. Proper leveling is essential for accuracy.

5.4 Připojení napájení

Connect the provided power cord to the balance and a suitable power outlet (100-240V). Alternatively, the balance can be powered by a 9V battery for portable use.

5.5 Doba zahřívání

After connecting power, allow the balance to warm up for at least 30 minutes before taking measurements. This stabilizes the internal components and ensures optimal accuracy.

6. Návod k obsluze

6.1 Zapnutí/vypnutí

  • Stiskněte tlačítko ON/OFF tlačítko pro zapnutí váhy.
  • Stiskněte a podržte ON/OFF tlačítko pro vypnutí váhy.

6.2 Základní vážení

To perform a basic weight measurement:

  1. Ensure the balance is level and has completed its warm-up.
  2. Stiskněte tlačítko TARE/NULA tlačítko pro nastavení zobrazení na 0.000 g.
  3. Opatrně umístěte vážený předmět do středu vážicí misky.
  4. Wait for the reading to stabilize. The stable weight will be displayed.

6.1 video: Demonstration of basic weighing, unit conversion, and percentage function on the U.S. Solid Lab Analytical Balance.

6.3 Funkce tárování

The tare function allows you to subtract the weight of a container, so only the net weight of the contents is measured.

  1. Na misku váhy položte prázdnou nádobu.
  2. Stiskněte tlačítko TARE/NULA button. The display will reset to 0.000g.
  3. Vložte vážený materiál do nádoby. Na displeji se zobrazí čistá hmotnost materiálu.

6.4 Převod jednotek

The balance supports multiple weighing units. To change the unit:

  • Stiskněte tlačítko JEDNOTKA button repeatedly to cycle through available units: g (grams), ct (carats), oz (ounces), ozt (troy ounces), lb (pounds), gn (grains), t (metric tons), dwt (pennyweights), kg (kilograms), dr (drams), tlT (Taiwan Tael), N (Newtons), and % (percentagE).
  • Vybraná jednotka se zobrazí na obrazovce.
Image showing the various measuring units available on the balance.
Obrázek 6.1: Available Measuring Units and Piece Counting options.

6.5 Parts Counting (PCS)

This function allows you to count a large number of identical items by weight.

  1. Place a known quantity of items (e.g., 10, 20, 50, 100, 150, 200, 250, 500) on the pan.
  2. Stiskněte tlačítko PCS tlačítko.
  3. Select the corresponding sample quantity using the JEDNOTKA tlačítko.
  4. Stiskněte PCS again to confirm. The balance is now in counting mode.
  5. Add more items, and the display will show the total count.

6.6 procenttage Vážení

Procentatage function is useful for formulation and quality control, allowing you to measure a sample's weight as a percentage referenční hmotnosti.

  1. Umístěte odkaz sample na misku váhy.
  2. Stiskněte tlačítko JEDNOTKA button until '%' is displayed. The reference weight will be set as 100%.
  3. Odstraňte odkaz sample.
  4. Place subsequent samples on the pan. Their weight will be displayed as a percentage referenční hmotnosti.

6.7 Kalibrace

Regular calibration ensures the accuracy of your balance. It is recommended to calibrate the balance upon initial setup, after relocation, or if you suspect inaccurate readings.

  1. Ensure the weighing pan is empty and the balance is stable.
  2. Stiskněte a podržte CAL tlačítko, dokud se na displeji nezobrazí „CAL“.
  3. The display will then show the required calibration weight (e.g., 500.000g).
  4. Carefully place the specified calibration weight (provided) onto the center of the weighing pan.
  5. Wait for the balance to automatically calibrate. Once complete, the display will show 'PASS' or return to normal weighing mode.
  6. Odstraňte kalibrační závaží.
Image showing the control panel buttons including ON/OFF, UNIT, PCS, CAL, and TARE/ZERO.
Obrázek 6.2: Control Panel Buttons. Key functions include ON/OFF, UNIT, PCS (Parts Counting), CAL (Calibration), and TARE/ZERO.

7. Údržba

7.1 Čištění

To maintain accuracy and prolong the life of your balance:

  • Před čištěním vždy odpojte napájení.
  • Wipe the exterior of the balance with a soft, damp tkanina. V případě potřeby použijte jemný prací prostředek.
  • Do not use abrasive cleaners, solvents, or immerse the balance in water.
  • Clean the stainless steel weighing pan and the windshield regularly to prevent accumulation of dust or residues.

7.2 Skladování

When not in use, store the balance in a clean, dry environment, away from extreme temperatures and humidity. Keep the windshield in place to protect the weighing mechanism.

8. Řešení problémů

This section addresses common issues you might encounter with your analytical balance.

ProblémMožná příčinaŘešení
Nepřesné nebo kolísavé hodnoty
  • Rovnováha není v rovině
  • Vibrations or drafts
  • Temperature changes
  • Vyžaduje kalibraci
  • Přetížení
  • Adjust leveling feet (Figure 5.1)
  • Relocate balance to a stable environment
  • Allow balance to acclimate to room temperature
  • Perform calibration (Section 6.7)
  • Ensure weight is within 500g capacity
Display shows 'Err' or other error codesVnitřní porucha nebo přetížení.
  • Remove all items from the pan.
  • Turn off the balance, wait 30 seconds, then turn it back on.
  • Pokud chyba přetrvává, kontaktujte zákaznickou podporu.
Balance does not power on
  • Žádné napájení
  • Faulty power adapter/battery
  • Zkontrolujte připojení napájecího kabelu a zásuvku.
  • Replace 9V battery if using battery power.
  • Test power adapter with another device if possible.
Cannot detect small weight changes (e.g., 10-15mg)
  • Environmental interference (drafts, vibrations)
  • Balance not properly calibrated
  • Warm-up time insufficient
  • Ensure stable environment, use windshield.
  • Proveďte kalibraci (kapitola 6.7).
  • Allow sufficient warm-up time (Section 5.5).

9. Specifikace

Technical specifications for the U.S. Solid 500g x 0.001g Lab Analytical Balance.

FunkceDetail
ZnačkaUS Solid
Číslo modeluUSS-DBS64-500
Hmotnostní limit500 gramů
Přesnost čtení0.001 Grams (1mg)
Typ zobrazeníLCD (podsvícený)
Typ váhyVáha rovnováhy
Měrné jednotkyg, ct, oz, ozt, lb, gn, t, dwt, kg, dr, tlT, N, %
NapájeníAC (100-240V) / DC (9V Battery)
Typ materiáluAkrylonitril-butadien-styren (ABS), nerezová ocel
Hmotnost položky3.64 libry
Zahrnuté komponentyScale Body, Scale Pan, Power Cord, Calibration Weight, Manual

10. Záruka a podpora

U.S. Solid products are manufactured to high-quality standards. For information regarding product warranty, please refer to the warranty card included with your purchase or visit the official U.S. Solid webmísto.

For technical support, troubleshooting assistance, or to inquire about replacement parts, please contact U.S. Solid customer service. Contact details can typically be found on the product packaging or the official brand webmísto.